Veterinary VET Portable Patient Monitor ICU machine SPO2/NIBP/ECG/TEMP/PR/RESP
Veterinary VET Portable Patient Monitor ICU machine SPO2/NIBP/ECG/TEMP/PR/RESP
Model:CMS6000VET
1.Functions
Standard parameters: ECG, RESP, SpO2, PR, NIBP, dual-channel TEMP
1)ECG Heart rate(HR)
ECG waveform
Arrhythmia and ST-segment analysis
2)RESP Respiration rate(RR)
Respiration waveform
3)SpO2 Oxygen saturation(SpO2)
Plethysmogram(PLETH) waveform
Pulse rate(PR)
Bar graph
4)NIBP Systolic pressure(SYS), Diastolic pressure(DIA), Mean pressure(MEAN)
5)TEMP T1, T2, TD
It has abundant functions, such as audible and visual alarm, trend data storage and output, NIBP measurement, alarm event marking and drug concentration calculation, etc.2.Performance
?ECG
Lead mode: 3-lead or 5-lead
Lead selection: I, II, III, aVR, aVL, aVF, V
Waveform: 5-lead, 2-channel
3-lead, 1-channel
Gain: 2.5mm/mV, 5.0mm/mV, 10mm/mV, 20mm/mV, 40mm/mV
Scan speed: 12.5mm/s, 25 mm/s, 50 mm/s
HR:
Measurement and alarm range: 15~350bpm
Accuracy: ±1% or ±1bpm, whichever is greater
Alarm accuracy: ± 2 bpm
Resolution: 1 bpm
ST-segment monitoring:
Measurement and alarm range: -2.0mV~+2.0mV
Accuracy: -0.8mV~+0.8mV ±0.04mV or ±10%, whichever is greater
Other range: unspecified
Arrhythmia analysis: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2,
PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC
Pacemaker: yes
?RESPIRATION
Method: R-F(RA-LL) Impedance
Respiration rate:
Measurement and alarm range: 0~150rpm
Resolution: 1 rpm
Measurement accuracy: ±2 rpm
Alarm accuracy: ±3 rpm
Apnea alarm: 10~40s
Scan speed: 6.25 mm/s, 12.5 mm/s, 25 mm/s
?NIBP
Method: Oscillometry
Mode: Manual/AUTO/STAT
Measurement interval in AUTO mode: 1/2/3/4/5/10/15/30/60/90/120/240/480/960 minutes
Measurement period in STAT mode: 5 minutes
Measurement and alarm range: 10 ~ 270 mmHg
Resolution: 1 mmHg
Cuff pressure accuracy: ±3 mmHg
Measurement accuracy:
Maximal mean error: ±5 mmHg
Maximal standard deviation: 8mmHg
Over-pressure protection:
Other mode: 297±3 mmHg
Dog mode: 240±3 mmHg
Cat mode: 147±3 mmHg
?SpO2
Measurement and alarm range: 0 ~ 100%
Resolution: 1%
Measurement accuracy: 70%~100%: ±2%;
0%~69%: unspecified
?PR
Measurement and alarm range: 30 ~ 250 bpm
Measurement accuracy: ±2 bpm or ±2%, whichever is greater
?TEMP
Channel: dual-channel
Measurement and alarm range: 0 ~ 50?
Resolution: 0.1 ?
Accuracy: ±0.1 ?
?Power supply: 100-240V?,50/60Hz
?Safety classification: Class I, type CF defibrillation-proof applied part
3.Accessories
1)Digital BD7 veterinary tongue-clip SpO2 sensor (1M)
2)General digital SpO2 sensor extension cable (2M)
3)Neonatal capsule single tube repeatable cuff (6-11cm)
4)NIBP extension tube
5)5-lead ECG lead cable (American-Standard)
6)ECG veterinary clip (*5)
7)Transvaginal TEMP probe
8)Power adapter
9)Power cord
10)User Manual4.Physical characteristic?
Dimension: 235 mm(L) ? 141 mm(W) ? 222 mm(H)
Weight: 2.5 Kg